Journal of the Vacuum Society of Japan
Online ISSN : 1882-4749
Print ISSN : 1882-2398
ISSN-L : 1882-2398
Regular Article
Attempts to Improve the Sensitivity and the Energy Resolution of an Analyzer for Auger Photoelectron Coincidence Spectroscopy and Electron Ion Coincidence Spectroscopy
Sadanori ARAERui KANEMURAKenta HIRAGAYosuke OGASHIWAKohtaro YANASENoritsugu KANAYAMAShinya OHNOTakuhiro KAKIUCHIKazuhiko MASEKoji K. OKUDAIRAMakoto OKUSAWAMasatoshi TANAKA
Author information
JOURNAL FREE ACCESS

2013 Volume 56 Issue 12 Pages 507-510

Details
Abstract

  We constructed two types of coincidence analyzers those can be used for Auger photoelectron coincidence spectroscopy and electron ion coincidence spectroscopy. The first one is designed for high electron energy resolution measurements, and consists of a double-pass coaxially symmetric mirror electron energy analyzer (DP-ASMA), a double-pass cylindrical mirror electron energy analyzer (DP-CMA), a time-of-flight ion mass spectrometer (TOF-MS), a magnetic shield, a conflat flange, an xyz stage, and a tilt adjustment mechanism. The second one is dedicated for high sensitivity measurements, and involves an ASMA with a large pinhole, DP-CMA with a large pinhole, and a TOF-MS.

Content from these authors
© 2013 The Vacuum Society of Japan
Previous article Next article
feedback
Top